beautypg.com

Creating a track view – Teledyne LeCroy WavePro 7 Zi_Zi-A User Manual

Page 111

background image

Getting Started Manual

922137-00 Rev A

103

Note: Turning off a trace for which trend data is being collected resets the
trend. If it is necessary to continue data collection for the trend, first create
a zoom trace of the channel trace (before turning off the channel trace).

Creating a Track View

1. In the menu bar, touch Measure → Measure Setup....

2. Touch a parameter tab Px and its corresponding dialog is shown.

3. Now, on the dialog, touch inside the Source1 data entry field and

select an input waveform from the pop-up menu.

4. Touch inside the Measure data entry field and select a parameter

from the pop-up menu.

5. Touch the Track button at the bottom of the dialog.

6. Now, from the Math selection for Track menu, select a math

function location (Fx) on which to store the Track display.

The Track is then shown along with its respective trace descriptor
label for the selected math function.