WavePro 7 Zi/Zi-A Oscilloscopes

142

922137-00 Rev A

Trigger Enabled - This function can be used as a gating function to
trigger another instrument when the oscilloscope is ready.

Trigger Out - This function can be used to trigger an external
oscilloscope.

Pass/Fail - Allows you to set a pulse duration (varied based on your
specific model, refer to datasheet specifications at
teledynelecroy.com for details); generates a pulse when pass/fail
testing is active and conditions are met.

Fast Edge - Some models provide this signal output used to deskew
multiple channels. The Fast Edge connection varies based on model
and is either SMA or BNC.

C

ALIBRATION

O

UTPUT

S

IGNALS

A calibration signal can be output from the Probe Calibration Hook on the
front of the oscilloscope. The following Calibration Output Signals are
available for use:

Square Wave - You can set a Frequency value (varied based on your
specific model, refer to datasheet specifications at
teledynelecroy.com for details).

DC Level - This is a reference level.

Off - This function turns off the Auxiliary Output signal.