WavePro 7 Zi/Zi-A Oscilloscopes

4

922137-00 Rev A

The instrument also has internal fan control circuitry that regulates the fan
speed based on the ambient temperature. This is performed automatically
after start-up.

Cleaning

Clean only the exterior of the oscilloscope using a damp, soft cloth. Do not
use harsh chemicals or abrasive elements. Under no circumstances
submerge the instrument or allow moisture to penetrate it. Avoid electric
shock by unplugging the power cord from the AC outlet before cleaning.

CAUTION. Do not attempt to clean internal parts. Refer to qualified
service personnel.

Calibration

The oscilloscope is calibrated at the factory prior to being shipped. The
recommended calibration interval is one year. Calibration should be
performed by qualified personnel only.

The oscilloscope software includes automatic and manual calibration
functions.

CAUTION. It is required that all inputs be removed from the
oscilloscope prior to performing a manual calibration.

Schedule an annual factory calibration as part of your regular maintenance.
Extended warranty, calibration, and upgrade plans are available for
purchase. Contact your Teledyne LeCroy sales representative or
[email protected] to purchase a service plan.

Power

AC Power Source

100-240 VAC (+/-10%) at 50/60 Hz (+/-5%); 100-120 VAC (+/-10%) at
50/60/400 Hz (+/-5%); Installation Category: 300V CAT II

Manual voltage selection is not required because the instrument
automatically adapts to line voltage.