WavePro 7 Zi/Zi-A Oscilloscopes

46

922137-00 Rev A

and total available memory to determine the actual number of samples or
segments, and time or points.

Setting up Sequence Mode (Adjacent)

1. Touch Timebase → Horizontal Setup... on the menu bar.

2. Click the Sequence tab.

3. Under Acquisition Settings, touch inside the Num Segments data

entry control and provide a value using your preferred input control
method. Additional information on using the touch screen controls
can be found in the Dialog Area (on page 33).

Note: The number of segments you choose to display can be less
than the total number of segments in the waveform. For example,
in the pop-up images above, the number of display segments is 10,
but the total number of segments entered in the timebase dialog's
Num Segments control is 100.

4. Touch the Enable Timeout checkbox.

5. Touch inside the Timeout data entry control and provide a timeout

value.

Note: Use the sequence mode timeout to automatically interrupt
the sequence acquisition if the timeout value is exceeded without a
valid trigger. The timeout period accounts for instances when a
Num Segments miscount occurs for some reason and the
oscilloscope waits indefinitely for an unforthcoming segment.