2.3.2 MOUNTING PROCEDURE

For single and multiple case mounting the instruments are furnished with a trim collar (mounting
frame). Figure 2-2 illustrates the installation and use of the trim collar (mounting frame). Trim col-
lars (mounting frames) are available in various sizes and are supplied to conform with the particu-
lar panel cut-out.

NOTE

Mounting brackets and trim collars (mounting frames) are packaged

separately. Check the shipment carefully to prevent loss of mounting

hardware.

To install single or multiple mounted instruments in a prepared panel cut-out, proceed as follows.

1.

Remove the through-case shipping bolt.

2.

Slip the trim collar (mounting frame) over the rear of the case and slide it forward to the
front of the case.

3.

Slide the instrument case through the panel opening.

a) Single mounting case - support the weight of the case and attach the top and bot-
tom mounting brackets. Tighten the bracket screws.

b) Multiple mounted cases - spacer bars and self-adhesive pads must be used
between the cases, as shown in Figure 2-3. Start the installation from the right
(when facing the panel), installing the spacers as each case is added. Also, as
each case is positioned in place, install and tighten the top and bottom mounting
brackets. Each case must be tight against the previous case.

NOTE

Spacers are not required on the outside of the right and left cases.

53IT5100 Indicator/Totalizer

2-2