Previous test, Prime, Print batch worklist – Luminex 100 IS User Manual, Version 2.3 User Manual
Page 77: Print report, Recalc, Replay batch, Report raw fluorescence, Resume, Sample probe down, Sanitize

x
MAP Technology
Using Luminex 100 IS 2.3 Software
PN 89-00002-00-071 Rev. C
5 - 35
Previous Test
This command is found in the Analysis window. Use this command
to view the previously run test in the batch.
Prime
Removes air from the system’s fluidic pathways by drawing
Luminex xMAP Sheath Fluid from the sheath fluid container. You do
not need to supply solution in a plate. Priming takes approximately
one minute. This command should also be used as part of the daily
startup routine.
Print Batch Worklist
Prints the status of each of the commands in the command list, as
well as the microtiter plate graphic.
Print Report
Prints the data interpretation report including the standards, controls,
graph of standards, and Samples data.
Recalc
Recalculates an affected test after a change is made to lot
information. There are two options: Manual recalculate and auto. If
the auto checkbox is selected, the Recalc button (manually
recalculate) is grayed out. In the auto selection mode, changes
automatically trigger the system to reanalyze the affected test. If you
are not using the auto mode, the ReCalc button is available after you
make a change such as changing a formula or lot, or invalidating or
validating a standard or control.
Replay Batch
Replays batch sample acquisition. All batch commands are replayed
allowing for batch acquisition viewing exactly as in real-time. You
cannot replay New Advanced Batches. When running a Replay
Batch command, the original acquisition data is not altered. A new
file is created. The most common use is for system demonstration
and training.
Report Raw Fluorescence
Enables the median fluorescence intensity (MFI) to appear on the
Analyte Report.
Resume
Resumes the process that was paused.
Sample Probe Down
Raises and lowers the sample probe to adjust the probe height. This
adjustment is recommended when using different microtiter plates or
when changing sample probes.
Sanitize
Uses the Luminex XYP reservoir due to volume requirement. The
sanitize command performs a similar function as the alcohol flush