Luminex sheath delivery system, Waste fluid container, Optical – Luminex 100 IS User Manual, Version 2.3 User Manual
Page 38: Xmap technology reagents, Optical -12 xmap technology reagents -12

Luminex 100 IS User Manual Version 2.3
x
MAP Technology
3 - 12
PN 89-00002-00-071 Rev. C
Luminex Sheath
Delivery System
For proper operation, place the Luminex SD system at the same level
as the base of the Luminex XYP instrument. Do not put it on top of
the Luminex 100 analyzer.
Waste Fluid
Container
The waste fluid container receives waste from the system. The waste
container should not be placed on top of the instrument.
Optical
The optical system consists of the optical assembly and the excitation
lasers. The optical assemblies do not require manual adjustment by
the user.
xMAP Technology
Reagents
The xMAP technology reagent system consists of classification
calibration microspheres, reporter calibration microspheres,
classification control microspheres, and reporter control
microspheres.
Note:
If you are not using the SD
system, sheath fluid levels must
be monitored manually. Check the
sheath fluid level before starting a
run or procedure.
Warning:
If biological samples have been tested with the
system, use your standard laboratory safety practices.
Caution:
Waste levels must be manually monitored. Do not
allow the waste container to overflow!
.