Luminex 100 IS User Manual Version 2.3
x
MAP Technology
5 - 26
PN 89-00002-00-071 Rev. C
Figure 5-18. Non-Competitive Assay
Examples of out-of-range sample labeling for competitive assays are
shown in Table 5-7. Figure 5-19 shows what a standard curve for a
competitive assay should look like.
Table 5-6. Non-Competitive Out-Of-Range Labeling
Condition
Concentration Label
MFI of Standard Referenced
Left of Curve
< min concentration standard
Lowest
Right of Curve
> max concentration standard
Highest
Table 5-7. Competitive Out-Of-Range Labeling
Condition
Concentration Label
MFI of Standard Referenced
Left of Curve
< min concentration standard
Highest
Right of Curve
> max concentration standard
Lowest