Luminex xyp instrument general, Luminex sd system general, Pc specifications – Luminex 100 IS User Manual, Version 2.3 User Manual
Page 32

Luminex 100 IS User Manual Version 2.3
x
MAP Technology
3 - 6
PN 89-00002-00-071 Rev. C
Luminex XYP
Instrument General
•
Ambient temperature: 15°C to 30°C (59°F to 86°F)
•
Humidity: 20% to 80%, noncondensing
•
Altitude: operation up to 2400 m (7874 ft) above mean sea level
•
Physical dimensions: 44 cm (17.25 inches) W x 60 cm (23.5
inches) D x 8 cm (3 inches) H
•
Weight: 15 kg (33 lbs.)
•
UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1
•
Pollution degree: UL Pollution Degree 2, as defined in Section
3.7.3.2 of UL 61010A-1
•
Heater operating range: 35°C to 55°C (95°F to 131°F) with
tolerance 0°C to +2°C
•
Input voltage range: 100-240 V~ ± 10%, 1.8 Amps, 47-63 Hz
•
AC inlet fuse: 3 A, 250 V~, fast acting
Luminex SD System
General
•
Ambient temperature: 15°C to 30°C (59° to 86°F)
•
Humidity: 20% to 80%, noncondensing
•
Altitude: designed to operate at up to 2400m (7874 feet) above
mean sea level
•
Physical dimensions: 20 cm (8 inches) W x 30 cm (11.75 inches)
D x 24.75 cm (9.75 inches) H
•
Weight: 9 kg (20 lbs)
•
UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1
•
Pollution degree: UL Pollution Degree 2, as defined in Section
3.7.3.2 of UL 61010A-1
•
Input voltage range: 100-240 V~ ± 10%, 0.4 Amps, 47-63 Hz
•
AC inlet fuse: 2 Amp, 250 V~, time lag
PC Specifications
For systems using a PC, A Dell OptiPlex GX280 or Dell Optiplex
GX520 (or newer PC) is shipped with the Luminex 200 system. For
systems using a laptop, a Dell D610 Notebook is shipped with the
system. Microsoft
®
Windows
®
XP is pre-installed on the computers.
The power requirements are 115-230 V~, 6 Amps, 50-60 Hz
For updated information regarding the PC, notebook, or operating
system, go to http://www.luminexcorp.com, then click on the
Support link to open the FAQ list.