beautypg.com

Capacity – Luminex 100 IS Version 2.1 User Manual

Page 9

background image

x

MAP Technology

Luminex xMAP Technology

PN 89-00002-00-062 Rev. A

3

Capacity

The following specifications reflect minimum capacity values:

10 GB hard drive

Store data for up to 240,000 test results

1.4 MB 3½” inch floppy disk

100 MB read/write CD ROM

Analyze multiple 96-well plates per batch

Analyze multiple assay templates per plate

Distinguish a minimum of 1 to a maximum of 100 unique xMAP
microsphere sets in a single sample

Detect and distinguish surface reporter fluorescence emissions at
575 nm on the surface of 1 - 100 unique xMAP microsphere sets
in a single sample

Reporter channel range: 3.5 decades of detection

Collect from 10 to 100,000 events per xMAP microsphere set
specified

2.5 x 10

6

microspheres/mL maximum bead concentration

Sample uptake volumes from 20 to 200

µL

Sample core: 15-20 µm core at 1 µL/sec. sample injection rate

Sample timeout limit from 1 to 400 seconds

Automatic sampling from a 96-well plate

Start sampling from any well position

Sheath container and waste container hold enough volume to run
up to two 96-well plates between refills

96-well microtiter plates types:

Must be compatible with XYP instrument plate holder: Flat
bottom, conical, round, filter bottom, half-well plates with
overall height no more than 0.75 inches (19 mm). Use black
opaque plates if possible to reduce photobleaching of beads.

Microtiter plates with 96 wells must be compatible with XYP
instrument heater block temperature from 35°C to 55°C
(tolerance 0°C to ± 2°C) (95°F to 131°F) when performing
heated assays and using the heater block