Performance specifications, Speed, Accuracy and precision – Luminex 100 IS Version 2.1 User Manual
Page 8: Sensitivity

Luminex 100 IS Developer Guide to xMAP Technology Version 2.1
x
MAP Technology
2
PN 89-00002-00-062 Rev. A
Performance
Specifications
Speed
•
Speed: greater than or equal to 933 MHz Pentium III processor
with 256 MB RAM
•
USB cabling for fast data transfer
•
Automatic transfer of assay templates and new reagent
information into the system using a 3½” inch floppy disk or large
capacity read/write CD
•
Installation < 4 hours
•
System calibration < 10 minutes
•
System controls < 10 minutes
•
Barcode reader entry of sample IDs
•
Automatic post analysis
•
Analyze one 96-well plate/hour depending on manufacturer’s kit
•
Up to 100 xMAP microsphere sets per sample
•
Sheath flowrate 90
µL/sec. ± 5 µL
•
Sample injection rate into detector area 1
µL/sec. ± 0.05 µL
•
System warmup 30 minutes: Systems that remain inactive for at
least four hours will require a warm-up to restart the lasers. After
acquiring sample, running system calibrators, running system
controls, and warming up the instrument, the system resets the
four-hour internal clock.
Accuracy and Precision
•
Sample uptake volume accuracy ± 5%
•
Classification of xMAP microspheres > 80.0%
•
Misclassification of xMAP microspheres < 0.5%
•
Temperature control 0°C to ± 2°C of target
•
Internal sample carry over < 0.9%
•
Soluble background fluorescence emission at 575 nm
automatically subtracted from fluorescence intensity value
Sensitivity
•
Detect 1000 fluorochromes Phycoerythrin (PE) per microsphere
•
Reporter channel dynamic range: 3.5 decades