Luminex 100 IS Version 2.1 User Manual

Page 38

background image

Luminex 100 IS Developer Guide to xMAP Technology Version 2.1

x

MAP Technology

32

PN 89-00002-00-062 Rev. A

3. Select Quantitative for the template type.

4. Enter the Sample Vol (µL) for this template. This is the amount

of sample that will be tested. Make the volume in the wells at
least 25 µL greater than the sample volume.

5. Enter a Sample Timeout (sec) for this template. This is the

maximum time in seconds that the analyzer will spend acquiring
each sample.

6. Enter the Low Limit and High Limit for the doublet

discriminator gate.

7. Click Next. The next Template Setup Wizard dialog box

appears.

Figure 17. Opening Standard

8. Click Apply Standard/Control to add a standards product to

your template. The Open Reagent Product dialog box appears:

Note:

Templates have a

default sample timeout of 25
seconds. The timeout begins
when the stream begins. If the
system does not acquire the
expected number of events in
the specified time, the system
processes the next sample.

Apply Standard/Control
button.