Recommended Additional
Equipment
Uninterruptible
Power Supply
(UPS)
Luminex highly recommends using an uninterruptible power supply
(UPS) to protect your system from power outages. Choose one that
can provide 1050 Watts for at least 45 minutes. The UPS should be
UL listed, CSA certified, and CE marked when used internationally.
Surge Protector
If you do not use a UPS, use a surge protector. Choose a protector
that meets your needs. Factors to consider include electrical
environment, endurance, suppressed voltage rating, and method of
protection. It should have six outlets, rated at least 1500 Watts, and
be UL listed, CSA certified, CE marked for nondomestic use when
used internationally.
Printer
Printer, HP LaserJet 2300 or available equivalent
Barcode Labels
Use the Code 128 barcode label type when scanning barcode labels
into the system as patient identities.
Vortex
VWR product number 58816-121: Speed range 0-3200 rpm or
equivalent
Bath Sonicator
Cole-Parmer
®
product number 08849-00: Operating frequency 55
kHz or equivalent
System Overview
The system consists of three subsystems: electronic, fluidic, and
optical. The following section describes the user-accessible
components of each subsystem.
Electronics
Power Input
Module
The power input modules contain the on/off switch and fuses.
Communications
Ports(SB9-PIN)
The communications port connects the Luminex 100 analyzer or the
Luminex XYP instrument to the computer, and the Luminex SD
system to the Luminex 100 analyzer.
Luminex 100
Analyzer
Ventilation Filter
Located on the bottom of the Luminex 100 analyzer, the filter must
be checked and cleaned as necessary. For proper ventilation, do not
obstruct the area below and allow at least two inches (5 cm) of
clearance around the Luminex 100 analyzer.