Luminex 100 IS User Manual Version 2.3
x
MAP Technology
B - 4
PN 89-00002-00-072 Rev. C
Figure B-3 Install Alignment Guide
7. Ensure that the power switches on the analyzer and XYP
instrument are in the off position.
8. Attach the power cord to the power input module of the Luminex
XYP instrument and attach the serial cable to the Luminex XYP
instrument. Do not plug the Luminex XYP instrument into the
power outlet.
9. Attach the power cord to the input module of the Luminex 100
analyzer, then attach the USB cable (PN 85-10011-00-046) to P1
on the analyzer. Do not plug the power cord into the power
outlet. See Figure B-4.
10. Connect the Luminex 100 analyzer USB cable and the Luminex
XYP instrument serial cable to the PC. See items 12 and 14 in
Figure B-4.
(
Note:
The cable that connects the
analyzer to the PC is 5 feet long; the
cable that connects the SD to the
analyzer is 2.5 feet long.