Error states, System error messages, Error states -13 system error messages -13 – Luminex 100 IS Version 2.3 User Manual
Page 167

x
MAP Technology
Troubleshooting
PN 89-00002-00-072 Rev. C
7 - 13
Error States
The following error states display on the Status Bar on the Run Batch
tab when an error occurs during system operation.
System Error
Messages
There are two types of error messages: system error messages, and
sample error messages.
System error messages are displayed in three places:
• the Message Log on the Diagnostics Tab
• the message log located in the Windows message log folder
• the Errors tab
You are using incompatible
sheath fluid.
Use only Luminex sheath fluid
in the Luminex 100 analyzer.
Other fluids may damage your
analyzer.
Error Message
Possible problem
Solution
Disconnected
The software hasn’t made initial
connection.
On the toolbar, click Connect;
wait for the status to change.
The communication cables are
not connected.
Remove and reconnect the cable
connections.
Close the software application
and reopen it.
Close the software application,
then turn off the Luminex
system. Power on the system
and check the status.
Refill Sheath
Sheath fluid is low.
Refill the sheath container, then
click Resume.
Running: Sheath Empty
Sheath fluid is empty.
Replace or fill the sheath fluid
container with sheath fluid. Run
two Prime commands.