Luminex xyp instrument air intake filter, Luminex xyp instrument air intake filter -6, See figure 6-2 – Luminex 100 IS Version 2.3 User Manual
Page 146

Luminex 100 IS User Manual Version 2.3
x
MAP Technology
6 - 6
PN 89-00002-00-072 Rev. C
Figure 6-2 Grasping the Tubing
4. Remove the filter with one hand, and hold the tubing with the
other hand.
5. Connect a new filter to the tubing and position the filter inside
the panel.
6. Reattach the panel door to the unit.
7. Plug in and power on the Luminex 100 analyzer.
Luminex XYP Instrument
Air Intake Filter
To replace the XYP instrument air intake filter:
1. Disconnect the Luminex XYP instrument from ac power by
turning off the power switch on the rear of the Luminex XYP
instrument, then unplugging the Luminex XYP instrument power
cord from the wall source.
2. On the back of the Luminex XYP instrument, to the left side,
gently remove the screen from the Luminex XYP instrument
filter.
Note:
Do not remove the screws. See Figure 6-3.