Network installation advisory, Prepare system for first use – Luminex 100 IS Version 2.3 User Manual
Page 208

Luminex 100 IS User Manual Version 2.3
x
MAP Technology
B - 26
PN 89-00002-00-072 Rev. C
Verify Successful
Firmware Upgrade
1. On the PC desktop click
StartÖProgramsÖLuminexÖLXRÖLX Firmware. The LX
Firmware Update Tool dialog box opens. See Figure B-19.
2. An entry in the Movement Device text box should read:
Firmware up-to-date
.
3. Turn the power off to the SD, XYP, and LX100.
4. Reconnect the serial cable to the XYP.
5. Connect the SD interface cable to the SD and LX100.
6. Turn on the power to the SD, The XYP, and the LX100.
Network Installation
Advisory
This section is intended for your network IT representative regarding
computer name changes. Luminex does not support network
installations. However, if your installation requires that you change
the computer name, perform these steps:
To change the computer name on Luminex 100 IS 2.3 systems:
1. Change the computer name following the Windows process.
2. When you reboot the system an error message appears warning
about database errors.
3. Reinstall Luminex 100 IS 2.3 software using the Repair option.
After the reinstall the database is available; you do not need to
reboot.
Prepare System for
First Use
After you install the hardware and software, run the following
commands to prepare the system for first time use. Also, consider
running these commands when the system has been idle for extended
periods of time. Refer to Maintenance Procedures (page 5-93) for
command operation. Run the following commands in order:
•
2 Backflushes
•
2 Drains
•
2 Alcohol flushes
•
3 Washes with water
Note:
The Windows XP security
settings “Virus Protection” and
“Automatic Updates” alerts have
been turned off by default. Please
refer to your IT department when
connecting to the network. If you
have questions, please call
Luminex Technical Support.