To drain the system:
1. On the Maintenance tab, click Eject/Retract. Ensure that the
reservoir is empty, or insert a plate onto the plate holder.
2. Make sure that the correct location is selected next to the Drain
button. Click Drain. A confirmation dialog box opens.
3. Click OK. The Device Activity box on the Status Bar indicates
that the system is draining.
Run Self-
Diagnostics
To run self diagnostics:
1. On the Maintenance tab, click Self-Diag.
2. Click OK. The system processes the various self- diagnostic
tests. When tests are complete, the Status Bar changes from a
“Processing” command state to an “Idle” command state. The
self-diagnostic tests should take less than one minute to
complete.
If the self diagnosis fails, you can obtain detailed information
regarding the results of the self-diagnostic test. See the following
“View Self-Diagnostic Details” section.
To view details of the self-diagnostics test that passed or failed:
1. Click on the Diagnostics tab and view the Message log. If a
diagnostics test failed, an error message displays with a yellow
background.
2. Double-click the yellow row to see a detailed description. An
Errors dialog box opens showing a list of passed and failed self-
diagnostic tests. Click OK to close this dialog box.
Using the Cleanup
Utility
Use the Cleanup Utility to:
•
Perform a disk cleanup
•
Delete the Message Log Directory
•
Delete the Batch Directory
To display the Cleanup Utility dialog box:
On the Tools menu, click Cleanup. The Cleanup Utility dialog
box opens. See Figure 5-62.