Re-run or recover incomplete batch -61 – Luminex 100 IS Version 2.3 User Manual
Page 103

x
MAP Technology
Using Luminex 100 IS 2.3 Software
PN 89-00002-00-072 Rev. C
5 - 61
Figure 5-31. Processing Multiple Plates
After the first microtiter plate in a multiple plate batch has been
acquired, the system pauses and prompts you to insert the next plate.
To process multiple plates during a batch or multi-batch:
1. Create the batch or multi-batch. See “Create a New Batch” on
page 5-53 or “Create a Multi-Batch” on page 5-58.
2. Insert the first plate.
3. On the Run Batch tab, click Start Plate to begin processing the
batch. When the initial plate is through, the system pauses and
displays the Insert Next Plate message in red underneath the
sheath pressure gauge.
4. Click Eject, then remove the acquired plate and load the next
plate for processing.
5. Click Resume. The system resumes the acquisition process.
Re-Run or
Recover
Incomplete Batch
An incomplete batch can be caused by situations such as a power
failure, software failure, marking a batch for deletion, or clicking
Cancel All. Use this procedure to re-run or recover an incomplete
batch.