Luminex 100 IS Version 2.3 User Manual
Page 199

x
MAP Technology
Luminex 100 IS 2.3 Software Installation
PN 89-00002-00-072 Rev. C
B - 17
Figure B-16 Software Information Window
3. Run system calibration. See Calibration Procedures (page 5-46)
to complete the verification process by running the system
calibrators and controls.
Luminex 100 Version 1.7
with Windows 2000 to
Luminex 100 IS
Version 2.3
To upgrade to Luminex IS 2.3 software you perform these
procedures in the following order:
•
Optionally, archive your “My Sessions” folder
•
Remove LMAT software
•
Remove Luminex 100 Version 1.7 software
•
Install Luminex IS 2.3 software
•
Verify successful upgrade
Archive “My
Sessions” folder
To preserve your Luminex 100 Version 1.7 session data, archive
your “C:\My Sessions” folder. Copy the files to another folder or
save them to a diskette or CD. Luminex 100 IS Version 2.3 software
maintains session data in a database, not in the My Sessions folder.
Remove Luminex
LMAT Software
Remove the Luminex LMAT software using the Windows Control
Panel.
1. On the PC desktop select:
Start
Ö
Settings
Ö
Control Panel
.
The Control Panel opens on the desktop.
2. Double-click the Add/Remove Programs icon. The
Add/Remove Programs dialog box opens.
3. In the Currently installed programs list, select LMAT entry. The
entry expands to include the Change/Remove button. Click
Change/Remove.
Note:
During the LMAT
software removal, error dialog
boxes may open. They are of
no consequence to the
uninstall. Just click OK or Yes
in response and continue with
the uninstall.