Figure 27. Correction Factor

Application of the correction equation will be illustrated in the following example. Through analysis
of the process flow stream it is determined that the flow rate must be multiplied by 1.056 to output the
desired value.

The correction equation takes the following form:

mnew = 0 + (1.056 x mo) + (0 x mo

2

)+ (0 x mo

3

)

Therefore mnew = 1.056 x mo
Where

F1 = 0
F2 = 1.056
F3 = 0
F4 = 0

A more complicated situation would be where the multiplication factor varies with flow rate.
Figure 28 shows the variation of desired flow rate versus the measured flow rate.

Is the

correction

equation

enabled?

Apply correction

equation

Corrected flow rate

Apply other

corrections (see

Aux. Input)

Yes

C00247-2

Manipulate

analog outputs

and relays

Measured flow

rate

No

Measured Flow Rate

Desired

Flow Rate

0

10

20

30

40

50

0

20

40

60

C00248-1

Figure 28. Desired Flow Rate Versus Measured Flow Rate