Quick operation procedure – Fluid Components International GF90_GF92 User Manual
Page 36

FLUID COMPONENTS INTERNATIONAL LLC
http://www.fluidcomponents.com
GF Series Flow Meter Models GF90/GF92
32
Doc. No. 06EN003229 Rev. H
Quick Operation Procedure
Input power is the only thing needed to operate the instrument. If a modification to the factory setup is
needed, continue with this section.
To use Table 6, an assumption is, the password is enabled (this is the factory preset condition).
For proper operation of the flowmeter all the information in the Analog Output category needs to be
entered as a group, also all of the information in the Flow Rate Units category needs to be entered (do
not skip steps).
If mistakes in data entry are made, press the back arrow key to correct the mistake.
Table 6. Quick Operation Procedure
FUNCTION DESIRED*
KEYS TO PRESS**
COMMENTS
Main Menu
Press any key
To get from the normal display to the
main menu.
Main Menu
Press HOME then press 1
To get from other menus to the main
menu in most cases. In some
cases (when shown) press ENTR,
then HOME.
(1) Normal Operation
Press HOME then press 1
To get to the normal display.
Set Password
(1) Enable/Disable
Press HOME, 4, 3, 1, enter the user
password, then press ENTR, HOME
The factory recommends the
password be disabled before any
information is entered to spped up
the process.
Port 1: Mode
(1) 4-20 mA (2) 0-5 V
(3) 0-10 V (4) 1-5 V
Press HOME, 2, 1, 1, 1, press 1,2,
3, 4, or ENTR, then press 5, Y or N,
HOME
Press key which best matches the
analog input to the peripherals from
Jack JP6 of the Flow Transmitter.
Press ENTR only if previous values
are correct.
Port 1: Unit
(1) Flow (2) Temp
Press HOME, 2, 1, 1, 2, then 1 or 2,
then 5, Y or N, HOME
Press the key which matches the
condition ot be measured.
Port 1 : F.S.
Press HOME, 2, 1, 1, 3, then enter a
numeric value. Then press ENTR, 5,
Y or N, HOME
Enter the numeric full scale value
which is expected to occur. The
values shown on the second and
third lines of the display. Press
ENTR without inputting a numeric
value only if previous values are
correct.
Port 1: Zero
Press HOME, 2, 1, 1, 4, then enter a
numeric value. Then press ENTR, 5,
Y or N, HOME
Enter the numeric zero scale value
which is expected to occur. The
values shown on the second and
third lines of the display. Press
ENTR without inputting a numeric
value only if previous values are
correct.
Port 2: Mode
(1) 4-20 mA (2) 0-5 V
(3) 0-10 V (4) 1-5 V
Press HOME, 2, 1, 2, 1, press 1,2,
3, 4, or ENTR, then press 5, Y or N,
HOME
Press key which best matches the
analog input to the peripherals from
Jack JP6 of the Flow Transmitter.
Press ENTR only if previous values
are correct.
Port 2: Unit
(1) Flow (2) Temp
Press HOME, 2, 1, 2, 2, then 1 or 2,
then 5, Y or N, HOME
Press the key which matches the
condition ot be measured.
ANALOG OUTPUTS