Service hold version, Aquamanager and the system menu – YSI 5200A User Manual
Page 63
YSI 5200A
124
YSI 5200A
125
Press
1.
to toggle relay on.
Press
2.
to toggle relay off.
Press
3.
to save configuration and return to Run Screen.
Note:
Timer and control icon(s) remain on the Run Screen until the system is
-
reset. Example - the timer icon remains at the systemwide location until
the end of the timer cycle that was stopped when the relay test menu was
entered even though that relay is not energized. To reset systems after
exiting the Relay Test menu, power cycle the 5200A.
Service Hold
Menus → System → Service Hold
Pressing
at the Run Screen puts the 5200A into Service Hold. The Service
Hold time is user selectable (1-255 sec, min, or hours). Service Hold turns off any
active relays and disables any sensor, aux, and timer system relays from energizing
regardless of value or condition. 5200A alarm emails are not sent during service
hold. After the service hold time expires or is cancelled, sensor systems are reset
after 30 seconds. Timer systems are not reset.
Use
1.
▲
and
▼
to scroll and highlight submenus.
Configuring the 5200A
Press
2.
to select.
Enable and configure submenus.
3.
Press
4.
to save configuration and return to Run Screen.
Version
Menus → System →Version
The Version menu contains information about the 5200A including: display and
control board firmware code version, control board temperature, and AC and DC
voltage.
Notes:
Firmware version information is required when contacting YSI sup-
-
port.
The clock chip temperature range starts at 0
-
oC. Values < 0oC are not valid
numbers.
Firmware date and time compilation are displayed below the version
-
number.
The normal operating temperature range will be a few degrees above
-
ambient.
AC powered units will run at a higher temperature than DC powered
-
units.
- To view current AC and DC power values press
to return to the
Run Screen and then return to the menu. Power information does not
refresh when the Version menu is open.
AquaManager and the System Menu
AquaManager users can configure most 5200A system menus from a PC. After poll-
ing the 5200A, use “Systems Management” from the node window to view and upload
5200A configurations. System menus are configured at the Setup and Input tabs.
Configuring the 5200A