Non-networked (stand alone) 5200as, Configure aux, Probe name input type – YSI 5200A User Manual
Page 50
YSI 5200A
98
YSI 5200A
99
Non-Networked (Stand Alone) 5200As
Non-networked (stand alone) 5200As must have the termination relay enabled. In
addition, non-networked (stand alone) 5200As must be configured as a master in
the System Communication menu with no slave nodes enabled.
Configure AUX
Menus → System → Configure Aux 1-2
The 5200A has two auxiliary (Aux) inputs. Aux inputs can be configured as digital or
analog (0-1VDC, 0-5VDC, or 4-20 mA) inputs. Configure suffix, minimum (Min),
maximum (Max), and display format menus for the analog inputs. Go to Sensor
Setup Menus → Sensor Setup Aux 1-2 - page 133 to enable Aux channel and configure
control/alarm settings. Both digital and analog directions are on the following pages.
Submenu functionality is provided after the configuration directions.
Probe Name
Menus → System → Configure Aux →Aux 1-2→Probe Name
Use the alpha/numeric keypad to name aux inputs. Configured name(s) appear
throughout menus and at Run Screen - figure 4.6, page 101.
Note:
Aux names can also be configured at Menus → System → Name Devices/
-
Sensors.
Input Type
Menus → System → Configure Aux →Aux 1-2→Input Type
Select the input type for the device(s) wired to aux system(s). Aux Sensor Setup
(page 133) menu selections are based on the aux input type configured at the Sys-
tem menu.
Configuring the 5200A
Digital
Menus → System → Configure Aux →Aux 1-2→Input Type→Digital
Use the digital input when wiring an external device to control or alarm when the
contacts of the external switch are open or close. A common application would be
to detect high or low water levels using a float switch.
Notes:
Suffix, Min, Max, and Display Format are not available for digital in-
-
puts.
In the 5200A sensor log, aux digital “closed” is logged as Aux 0.00. Aux
-
digital “open” is logged as Aux 1.00.
Wire aux digital devices -
1.
page 49.
Enter Aux menu
2.
Menus → System → Configure Aux →Aux 1-2
Use
3.
▲
and
▼
to scroll and highlight submenus.
Press
4.
to select.
Configure digital input type and name.
5.
Press
6.
to save configuration and return to Run Screen.
Go to Sensor Setup Menus → Sensor Setup Aux 1-2 -
7.
page 133 to enable
and configure control/alarm on condition, control/alarm relay, and hold
off time.
Press
8.
to save configuration.
Configuring the 5200A