YSI 5200A User Manual
Page 46
YSI 5200A
90
YSI 5200A
91
address menu.
When DHCP is enabled, the serial to ethernet device acquires SMTP and
-
DNS addresses during initialization.
- AquaManager requires a static IP address; therefore, DHCP must be
disabled when using AquaManager. Consult the network administrator
to obtain a static IP address.
Power cycle the 5200A whenever DHCP is enabled.
-
IP Address
Menus → System → Communications → Ethernet → IP Address
The IP address can only be configured when DHCP is disabled.
Notes:
- AquaManager users should program the Gateway address either by force
serial port in the AquaManager Communication menu or locally at the
5200A. Poll node using RS232 communication to upload configuration
information including: static IP, port, subnet, and gateway addresses.
When DHCP is enabled, the IP address is not updated at the IP address
-
menu.
Power cycle the 5200A whenever a new IP address is configured.
-
Subnet Mask
Menus → System → Communications → Ethernet → Subnet Mask
A Subnet mask address can only be configured when DHCP is disabled. The subnet
mask number helps to define the relationship between the host (computers, rout-
ers, switches, etc.) and the rest of the network. Consult the network administrator
when configuring this menu.
Notes:
When DHCP is enabled, the Subnet Mask address is not updated at the
-
Subnet Mask menu.
- AquaManager users should program the Subnet mask either by force serial
port in the AquaManager Communication menu or locally at the 5200A.
Poll node using RS232 communication to upload configuration informa-
tion including: static IP, port, subnet, and gateway addresses.
Gateway
Menus → System → Communications → Ethernet → Gateway
The Gateway address can only be configured when DHCP is disabled. A gateway IP
refers to a device on a network which sends local network traffic to other networks.
Consult the network administrator when configuring this menu.
Configuring the 5200A
Notes:
When DHCP is enabled, the Gateway address is not updated at the
-
Gateway menu.
AquaManager users should program the Gateway address either by force
-
serial port in the AquaManager Communication menu or locally at the
5200A. Poll node using RS232 communication to upload configuration
information including: static IP, port, subnet, and gateway addresses.
Port
Menus → System → Communications → Ethernet → Port
The default port number for the ethernet module is 10001. Consult the network
administrator when configuring this menu.
Note:
- AquaManager users should program the port number either by force
serial port in the AquaManager Communication menu or locally at the
5200A. Poll node using RS232 communication to upload configuration
information including: static IP, port, subnet, and gateway addresses.
DNS Address
Menus → System → Communications → Ethernet → DNS
Domain Name Service (DNS) numbers are IP addresses that a workstation or server
uses to refer to specific servers that resolve domain names to IP addresses.
Notes:
When DHCP is enabled, the serial to ethernet device will acquire SMTP
-
and DNS addresses during initialization.
When DHCP is enabled, the DNS address is not updated at the IP ad-
-
dress menu.
Email Addresses
Menus → System → Communications → Ethernet → Email Address
→ Email Address
→ Email Address
→ CC/Email Address
Up to four email addresses can be configured for 5200A emails. At a minimum,
the first email address and the CC/email address must be configured. If the CC/
email address is not configured, 5200A alarm emails will not contain any alarm
information. The email ‘from’ address will be the email address entered in the first
Email Address menu.
Configuring the 5200A