Wire dc power and backup power – YSI 5200A User Manual
Page 23
YSI 5200A
44
YSI 5200A
45
D
o not c
onnec
t or disc
onnec
t IDC r
ibbon cable
when 5200A is po
w
er
ed
. S
er
ious damage can oc
cur
.
Figure 3.17
Complete other wiring to 5200A.
5.
Close front panel -
6.
page 40.
Test AC power.
7.
Test AC Power
Turn on main power. The enabled sensor values will display on the Run Screen after
the initial YSI splash screen. If applicable, install and configure DC power backup
page 45. Verify AC power at the Menus → System →Version menu - page 125.
See Power Fail Menu - page 117 for AC power event log information.
Wire DC Power and backup power
The 5200A may be powered by most 9.0-16.5 VDC source that provides 800mA of
current and is isolated from mains supply by double or reinforced insulation. The
DC power source is user-supplied and can come from a variety of choices including
lead acid, gel cell, or UPS (with 12VDC transformer) external batteries. DC power
wires are terminated at location B on the I/O Board - figure 3.18.
Notes:
The 5200A does not charge batteries. Quality assurance maintenance
-
procedures should be established if batteries are to be fully powered when
used as back-up power source.
Power supply voltage above 16.5 VDC may permanently damage the
-
5200A.
Installation and Wiring
When back up voltage falls below 9.0 volts, the 5200A ceases to operate
-
properly.
Some switching DC power supplies not supplied by YSI may result in
-
noisy readings.
Directions to Wire DC Power/DC Backup Power
WARNING: Disconnect external power to the unit before wiring.
Follow
all manufacturer’s safety and installation instructions when providing power to
the 5200A via a DC power supply.
AVERTISSEMENT : déconnectez l’alimentation externe de l’unité avant
d’effectuer le câblage. Suivez toutes les consignes de sécurité et d’installation du
fabricant lorsque vous fournissez de l’alimentation au modèle 5200A par un bloc
d’alimentation à courant continu.
Perform steps 1 - 3 of 5200A installation.
1.
Open front panel -
2.
page 39.
Feed DC power cable through drilled hole in rubber grommet to location
3.
B. The terminal strip is removable.
CAUTION
: Run high and low voltage cables through separate bulkhead and
conduit.
Terminate DC power wires to pins (-) and (+) according to manufacturer
4.
instructions and any applicable local electrical codes.
D
o not c
onnec
t or disc
onnec
t IDC r
ibbon cable
when 5200A is po
w
er
ed
. S
er
ious damage can oc
cur
.
Figure 3.18
Installation and Wiring