Softkey label
Function
Exit menu and return to Run Screen saving
any menu configuration changes.
Exit menu (to previous menu) without sav-
ing configuration changes.
Back to previous menu. Changes to configu-
rations are saved at return to Run Screen.
Prompt to take next action, i.e. enter a value
at calibration menus.
Save alpha and numeric configurations and
calibrations.
Disable/Enable menu option, i.e. buzzer.
Scroll left and right in alpha and numeric
keypad screens.
To acknowledge and save sensor calibration.
Prompt to confirm “Are you sure?” message
when performing certain functions, i.e. reset
to factory and clear data log menus.
Digital input control configuration from
Sensor Setup Menu.
Table 4.1
Run Screen
The Run Screen displays measurements and inputs from enabled sensors and con-
ditions - figure 4.2. At power up, the Run Screen is displayed immediately after
the YSI splash screen (the YSI splash screen appears for approximately 2 seconds).
Icons displayed at the Run Screen represent individual and systemwide conditions.
See Run Screen Icons - page 67.
Run Screen and sensor display formats are user selectable. See Display Format - page
107 and Sensor Setup Menus - page 127.
Notes:
The Run Screen must be displayed during normal operation. All alarm
-
functions (5200A alarm emails, energizing of enabled alarm relays,
alarm icons, and buzzer) are suspended when in 5200A instrument
menus. Alarm systems are reset when menus are exited to Run Screen.
Configuring the 5200A
- Always manually poll the 5200A after uploading configuration changes
and verify the uploaded information is retained. When configuration
changes are uploaded using AquaManager at the same time configuration
changes are being made at the instrument, the AquaManager uploads may
not be saved depending on the timing of keystrokes at the 5200A.
- The Run Screen must be displayed during normal operation. If the
5200A is not at the Run Screen during AquaManager uploading, pressing
at the 5200A will only save certain configuration changes.
- At power up, 5200A systems have a 30 second hold off allowing sensor
values to stabilize before enabling any control and/or alarm devices.
Figure 4.2
Run Screen Icons
Icons identify control, alarm, timer, power, service, and communication informa-
tion. Icons along the top of the Run Screen indicate 5200A systemwide conditions
- table 4.2. Icons displayed within a sensor window indicate condition(s) specific
to that system - table 4.3.
Configuring the 5200A