Firmware can only be uploaded to an instrument using a local RS232 con-
-
nection. See Downloader - page 109.
On a network, install and configure RS232 or the serial to ethernet device
-
(hardware and wiring) on the master node only.
Email Alarming
Email alarming can be configured in one of two ways:
- from a PC running the AutoPoll Program that is included with AquaMan-
ager software (PC must have internet connection). This is referred to as
AquaManager email.
via a serial to ethernet* device with TCP/IP connection at master 5200A.
-
This is referred to as 5200A email.
*Serial to ethernet connection requires access to a SMTP server accessible by the
local network.
RS232 Communication
Use Category 5e patch cable from RS232 port at location H - figure 3.29 of I/O
board to PC using RJ45 to DB9 adapter.
Directions to Wire Direct RS232 PC Connection
WARNING: Disconnect external power to the unit before wiring
.
AVERTISSEMENT : déconnectez l’alimentation externe de l’unité avant d’effectuer
un câblage quelconque.
Perform steps 1 - 4 of 5200A installation.
1.
Open front panel -
2.
page 39.
Feed direct connect cable through drilled hole of rubber grommet and
3.
to location H.
Run high and low voltage cables through separate bulkhead and con-
4.
duit.
Connect one end of communication cable to RJ45 at location
5.
H. Connect
other end to PC com port using RJ45 to DB9 adapter.
Installation and Wiring
D
o not c
onnec
t or disc
onnec
t IDC r
ibbon cable
when 5200A is po
w
er
ed
. S
er
ious damage can oc
cur
.
Figure 3.29
Complete other wiring to 5200A.
6.
Close front panel -
7.
page 40.
Test RS232 communication.
8.
Notes:
- When using a USB-Serial Adapter, verify that the port number assigned
during driver installation is the same port configured when using Aqua-
Manager and/or the Firmware Updater.
When making RS232 cable runs greater than 100 feet (30.5 m), use an RS232
-
Line Extender/Booster. Mount the booster in an area where it is protected
from moisture.
Proper PC com port number must be configured when using AquaManager
-
or Firmware Updater programs.
Test RS232 Communications
Apply power to 5200A. Insert node into AquaManager program. Poll node. If
wired correctly, AquaManager will get 5200A configuration and logs.
Ethernet (TCP/IP) Module Communication
The optional serial to ethernet device must be installed on the 5200A control board
when using TCP/IP for communication. Complete installation instructions are
provided with the optional serial to ethernet device.
Installation and Wiring