Node network – YSI 5200A User Manual
Page 47
YSI 5200A
92
YSI 5200A
93
Notes:
See
-
Send Test Email Msg - page 96 to verify proper configuration of eth-
ernet menus.
On a network, both master and slave alarm emails are sent to the config-
-
ured email addresses in the master.
Test the alarm email delivery by creating an alarm condition at the 5200A
-
and verify the email(s) is received.
The CC/Email address must be configured when using 5200A email alarm-
-
ing. Specific alarm system information will not be sent if the CC/Email
address menu is not configured with a valid email address.
Use AquaManager to easily configure email addresses using the PC key-
-
board. See AquaManager on line help.
- If Autopoller is being used for AquaManager email alarming on a TCP/
IP connection, it is strongly recommended that the 5200A email alarming
using the serial to ethernet device be disabled. When the ethernet device
sends alarm emails, all Autopoller SMS communication is suspended.
This could potentially delay any AquaManager alarm emails from being
sent.
To disable the serial to ethernet device from sending alarms, clear all four
-
email addresses from the Communication → Ethernet menus. Clear the
email addresses by using the “Spc” (space) key on the numeric keypad.
Save change at numeric keypad and press
to save configuration
and return to Run Screen.
Ethernet Module Error Message
At power up, the serial to ethernet device will generate an error message if it cannot
properly handshake with the 5200A hardware. If this error occurs, communication
(AquaManager and 5200A alarm emails) will not work via TCP/IP. The error could
be a result of incorrect serial to ethernet device installation or a faulty ethernet device.
The error message will appear in the event log.
Configuring the 5200A
SMS Messaging with Ethernet Module
SMS (Short Message Service) or text messaging is a service for sending messages to
your cell phone or mobile device. To use SMS messaging, enter the email address
in the Menus → System → Communications→ Ethernet →Email address. See Ap-
pendix 4 SMS - page 224 for a list of cellular and mobile companies that support
sending text messages.
Node Network
Menus → System → Communications → Network
A node network includes multiple 5200A and/or 5400 instruments. The devices
are physically hardwired together using RS485 protocol. Node networks are con-
figured with one master node and up to 31 slave nodes. The master continuously
queries enabled slaves on the network. The master is the access point for the entire
network. This includes AquaManager communication and sending of 5200A/5400
alarm emails. Slave nodes must be configured with a slave address.
Note:
- Multiple node networks communicating via direct connect (RS232) re-
quire separate PC COM ports for simultaneous AquaManager mapping
and autopolling.
Wire the 5200A network -
1.
page 52.
Enter Network menu
2.
Menus → System → Communications → Network
Press
3.
▲
and
▼
to scroll and highlight submenus.
For master node:
4.
Press
at master submenu.
For all slave nodes:
5.
Press
at slave submenu. Configure a slave ad-
dress.
Configuring the 5200A