YSI 5200A User Manual
Page 106
YSI 5200A
210
YSI 5200A
211
ORP(Enabled)
ORP (Enabled)
Alarms (Disabled)
Alarm Low (340 mv)
‐2000 to 2000*
Relay (Disabled)
Alarm Low (Disabled)
Instrument (Local Machine)
Relay (Relay 1)
�
Relay 1
Relay 2
Relay 3
Relay 4
Range Low (350mv)
341 to 2000*
Relay (Disabled)
Range Low (Disabled)
Instrument (Local Machine)
Relay (Relay 1)
�
Relay 1
Relay 2
Relay 3
Relay 4
Run Screen → Menus → Sensor Setup → ORP
Menu Maps
Menu Maps
Set Point (400 mv)
351 to 2000*
Range High (425 mv)
Relay (Disabled)
Range High (Disabled)
Instrument (Local Machine)
Relay (Relay 1)
T
Relay 1
Relay 2
Relay 3
Relay 4
Alarm High 2 (435 mv)
426 to 2000*
Relay (Disabled)
Alarm High (Disabled)
Instrument (Local Machine)
Relay (Relay 1)
T
Relay 1
Relay 2
Relay 3
Relay 4
pH (Enabled)
pH (Enabled)
Alarms (Disabled)
Alarm Low (8.10 pH)
0 to14.0*
Relay (Disabled)
Alarm Low (Disabled)
Instrument (Local Machine)
Relay (Relay 1)
�
Relay 1
Relay 2
Relay 3
Relay 4
Range Low (8.20 pH)
8.2 to 14.0*
Relay (Disabled)
Range Low (Disabled)
Instrument (Local Machine)
Relay (Relay 1)
�
Relay 1
Relay 2
Relay 3
Relay 4
Dose Up (5 Sec)
1‐255
Time Units
�
Sec
Min
Hours
Dose Up Wait (15 Min)
1‐255
Time Units
�
Sec
Min
Hours
Set Point (8.30 pH)
8.30 to 14*
Range High (8.40 pH)
8.40 to 14.0*
Run Screen → Menus → Sensor Setup → pH
continued next page