Chapter 5

– 52 –

PCI-1710 series User’s Manual

Advantech Co., Ltd.
www.advantech.com

Step 3: Adjust the registers until they fall between the input voltage

from the standard voltage reference and the receiving voltage
reflectected in the Manual A/D Calibration tab.

D/A channel Manual-Calibration

Step 1: Click the Manual D/A Calibration tab to show the D/A

channel manual calibration panel. Two D/A channels are
individually calibrated . Before calibrating, output desired
voltage from the D/A channels and measure it through an
external precision multimeter.

Step 2: For example, choose channel 0; select the Range and select

the wished output voltage code or value from the radio
buttons (Fig. 5-17 and Fig. 5-18).

Figure 5-17 & Figure 5-18: Selecting D/A Range and Choosing

Output Voltage