beautypg.com

2 pci-1711/1711l calibration, Figure 5-2: pci-1711/1711l vr assignment, Pci-1711/1711l calibration – B&B Electronics PCI-1711 - Manual User Manual

Page 49: Figure 5-2, Pci-1711/1711l vr assignment, Chapter 5

background image

Chapter 5

– 41 –

PCI-1710 series User’s Manual

Advantech Co., Ltd.
www.advantech.com

5.2 PCI-1711/1711L Calibration

Three calibration programs are included on the companion CD-ROM :

ADCAL.EXE

A/D calibration program

DACAL.EXE

D/A calibration program

(only for PCI-1711)

SELFCAL.EXE

D/A self-calibration program

(only for PCI-1711)

These calibration programs are designed only for the DOS environ-
ment. Access these programs from the default location:

C:\Program Files\Advantech\ADSAPI\Utility\PCI1711

VR Assignment

There are four variable resistors (VRs) on the PCI-1711 card and two
variable resistors (VRs) on the PCI-1711L card. These variable resistors
are to facilitate accurate adjustments for all A/D and D/A channels.
Please refer to the following figure for the VR positions.

Figure 5-2: PCI-1711/1711L VR assignment

The following list shows the function of each VR :

VR

Function

VR1

A/D bipolar offset adjustment

VR2

A/D full scale (gain) adjustment

VR3

D/A channel 0 full scale adjustment

(for PCI-1711 only)

VR4

D/A channel 1 full scale adjustment

(for PCI-1711 only)

PCI-1711 Series

CN1

VR1

VR2

VR3

VR4