beautypg.com

B&B Electronics PCI-1711 - Manual User Manual

Page 52

background image

Chapter 5

– 44 –

PCI-1710 series User’s Manual

Advantech Co., Ltd.
www.advantech.com

Figure 5-3: PCI-1716/1716L VR assignment

Calibration Utility

The calibration utility, AutoCali, provides four functions - auto A/D
calibration, auto D/A calibration, manual A/D calibration and manual
D/A calibration. The program helps the user to easily finish the calibra-
tion procedures automatically; however, the user can calibrate the PCI-
1716/1716L manually. Appendix E illustrated the standard calibration
procedures for your reference. If you want to calibrate the hardware in
your own way, these two sections will guide you. The following steps
will guide you through the PCI-1716/1716L software calibration.

Step 1: Access the calibration utility program AutoCali.exe from the

default location:

C:\Program Files\Advantech\ADSAPI\Utility\Auto Calibration

Note:
✎ If you installed the program to another directory, you can find this

program in the corresponding subfolders in your destination directory.

Step 2: Select PCI-1716/1716L in the ADSDAQ dialog box.

PCI-1716 Series

P1

SW1

VR1

TP4