beautypg.com

Figure 5-14: d/a calibration is finished – B&B Electronics PCI-1711 - Manual User Manual

Page 58

background image

Chapter 5

– 50 –

PCI-1710 series User’s Manual

Advantech Co., Ltd.
www.advantech.com

Step 12: Auto-calibration is finished (Fig. 5-14)

Figure 5-14: D/A Calibration is finished

A/D channel Manual-Calibration

Step 1: Click the Manual A/D Calibration tab to show the A/D

channel manual calibration panel. Before calibrating, acquire
the reference voltage from a precision standard voltage
reference. Go to the Range form, select a channel and the
target voltage range according to the input voltage value
from a precision standard voltage reference(Fig. 5-15).

Note:
✎ The input voltage value you selected from a precision standard

voltage reference needs to correspond with the one that the PCI-1716/
1716L can read.

✎ The input voltage will be analog code so the computer will convert the

voltage data into digitial code; therefore, the input voltage value you
selected from a precision standard voltage reference needs to
correspond with the one that the PCI-1716/1716L can read. For
example, if the input range is 0 ~ 5V, then input voltage should be
2.9992V not 3V.