Self A/D Calibration

We know, in most cases, it is difficult to find a good enough DC
voltage source for A/D calibration. We provide a self-adjusted A/D
calibration program “SELFCAL.EXE” to help solve this problem. The
steps of self-calibration are outlined as below:

1. Connect DA0_OUT to AI0.

2. Connect AGND to AI1, AI2, AI3, AI4 and AI5.

3. Run the SELFCAL.EXE program.

4. First calibrate the D/A channel. Adjust VR3 until the DA0_OUT

output voltage approaches +10V. Then press the SPACE key.

5. Next we will do the A/D calibration. Now the DA0_OUT output

voltage will be +9.995V, then adjust VR2 until the output codes from
the card’s AI0 focused on FFE (at least 70%) and adjust VR1 until
the output codes from the card’s AI1, AI2, AI3, AI4 and AI5
focused on 7FF (at least 70%).

6. Press the SPACE key to finish calibration procedures.

5.3 PCI-1716/1716L Calibration

A calibration utility, AutoCali, is included on the companion CD-ROM :

AutoCali.EXE

PCI-1716/1716L calibration utility

This calibration utility is designed for the Microsoft©Windows™
environment. Access this program from the default location:

C:\Program Files\Advantech\ADSAPI\Utility\Auto Calibration

VR Assignment

There is one variable resistor (VR1) on the PCI-1716/1716L to adjust
the accurate reference voltage on the PCI-1716/1716L. We have
provided a test point (See TP4 in Figure 5-3) for you to check the
reference voltage on board. Before you start to calibrate A/D and D/A
channels, please adjust VR1 until the reference voltage on TP4 has
reached +5.0000 V. Figure 5-3 shows the locations of VR1 and TP4.