beautypg.com

Zlinx i/o monitor, 8 zlinx i/o monitor – B&B Electronics ZXT9-IOA-KIT - Manual User Manual

Page 53

background image

Configuration & Operation

Manual Documentation Number: ZXTx-IO-x-4113m

47

To set Calibration for Analog Outputs:

1. Enter a value in the Reference text box.

2. Click the Set Current value button.

3. Supply power to the device.

4. Measure the output value and enter the measured value in the Acquired text box.

5. Values will be filled in automatically in the calibration field.

If using the Two Point method, repeat these steps for the second calibration point.

Fig 6-21 Analog Output Calibration

6.1.8 Zlinx I/O Monitor

From Zlinx Manager the user can choose the option Zlinx I/O Monitor. This option allows the monitoring of all Analog and
Digital I/O values (in V, mA, on/off; degrees Celsius for RTD) real time.

To start the Monitor feature:

1.

Start Zlinx Manager.

2.

Choose Zlinx I/O Monitor from the list.

3.

Zlinx I/O Monitor window will open.

4.

Choose the COM port the system is connected to.

5.

Click on the Start Monitoring button.