EN-68 7000 Series Indicators

SYMPTOM

PROBABLE CAUSE

REMEDY

ERROR 8.6 displayed.

The weight value cannot be displayed in the
current unit of measure because it exceeds
6 digits.

Reduce load on scale until weight value can be
displayed.

Use a more appropriate unit of measure.

------- displayed.

Busy message. Displayed during tare
setting, zero setting, printing

If this message persists, it usually indicates the
reading is not stable. Correct the instability.

--NO-- displayed.

The action is not allowed.

Do not attempt this operation.

Battery symbol flashing.

Batteries are discharged.

Replace batteries (T71P).

Charge batteries (when optional rechargeable
batteries are installed).

CAL E displayed.

Calibration value outside allowable limits.

Use correct calibration weights.

NO LOCK SW displayed.

Attempting to exit the menu with the Legal
for Trade setting ON and the security switch
OFF.

Set the security switch to the ON position, then exit
the menu. (See Section 6.1.)

REF WT ERROR displayed. Reference Weight too small. The weight on

the platform is too small to define a valid
reference weight.

Use a greater weight for the sample.

7.3

Service Information

If the troubleshooting section does not resolve your problem, contact an Authorized Ohaus Service Agent. For service assistance
in the United States, call toll-free 1-800-526-0659 between 8:00 AM and 5:00 PM Eastern Standard Time. An Ohaus Product
Service Specialist will be available to assist you. Outside the USA, pleas visit our website www.ohaus.com to locate the Ohaus
office nearest you.