7000 Series Indicators EN-19

CAL

CAL TEST

5LIN

lb

0

5LIN

lb

--C--

5LIN

lb

--C--

5LIN

lb

2.5

0.000

lb

5LIN

lb

DONE

5LIN

lb

--C--

tESt

kg

0

tESt

kg

10

tESt

kg

--T--

10.000

kg

0.001

kg

DIFFERENCE

tESt

kg

--T--

5LIN

lb

7.5

Place the specified 1/2 load calibration weight on the scale and press Yes.
The display shows --C--, followed by the 1/4 load calibration point.







Place the specified 1/4 load calibration weight on the scale and press Yes.
The display shows --C--, followed by the zero calibration point.





With no load on the scale, press Yes. The display shows --C--, then DONE and returns to
the current application mode.








3.3.5

Calibration Test

Calibration Test is used to compare a known calibration weight against the stored span calibration
data.

With no load on the scale, press Yes. The display shows the zero load calibration point.



With no load on the scale, press Yes. The display shows --T--, followed by the full load
calibration point.






Place the specified full load calibration weight on the scale and press Yes. The display shows --T--,
followed by difference between the calibration weight and the stored calibration data.






After 5 seconds, the display returns to the current application mode.