3 misc submenu, 1 display brightness, 2 ramp segments – American Magnetics 05100PS-430-601 Integrated Power Supply System User Manual
Page 79: 3 ramp rate time units
Rev. 5
61
Operation
Setup Submenu : Misc
3.10.3 Misc
Submenu
When the Misc submenu is selected, several miscellaneous parameters
may be viewed and/or changed.
3.10.3.1
Display Brightness
This picklist value controls display brightness. As shown above,
there are four brightness settings from which to choose (25%, 50%,
75% and 100%). The default setting is 100%.
3.10.3.2
Ramp Segments
The ramp segments value specifies the number of current ranges
which can be given unique ramp rate values. The default value is 1
unless preset by AMI to match a specific superconducting magnet.
When this value is 1, there is only one ramp rate for the Model 430
Programmer, used for the full available current range. For multiple
ramp rates, set the value to the number of ramp segments desired
(up to ten segments). See section 3.7.1 on page 35 for details
regarding the use of ramp rate segments.
3.10.3.3
Ramp Rate Time Units
This picklist value specifies the unit of time used to enter and the
display ramp rate. If Seconds is selected, ramp rate is entered in A/
s, kG/s or T/s; if Minutes is selected, ramp rate is entered in A/min,
kG/min or T/min. The selected unit value also applies to remote
interface commands. The default setting is Seconds.
+50.00 A — Display Brightness (%)
+0.50 Vs
25
50
75
100
+50.00 A — Ramp Segments (1-10)
+0.50 Vs
1
+50.00 A — Ramp Rate Time Units
+0.50 Vs
Seconds
Minutes