3 reset quench shift-key, 4 increment field shift-key, 5 field <> current shift-key – American Magnetics 05100PS-430-601 Integrated Power Supply System User Manual
Page 57: D 3.7.5, Operation
Rev. 5
39
Operation
Shift Key Commands : Reset Quench
If Voltage Limit becomes active while ramping, it will be indicated by a
reverse illumination character “V” for the status indicator.
Once the Voltage Limit function becomes active, the current, and therefore
field, will no longer be ramping linearly with time as the voltage available
to charge the magnet will be reduced as the total loop voltage will be
limited. As the IR drop of the leads increased with current, the voltage
available to charge the magnet will be reduced.
3.7.3
Reset Quench SHIFT-key
The
RESET QUENCH
SHIFT-key is used whenever a quench detection has
occurred and is being indicated on the display (example shown below).
When a quench detection has occurred, the Model 430 Programmer will
respond to no further input until the
RESET QUENCH
SHIFT-key is used,
or until the quench condition is cleared by a remote command. See Refer to
3.7.4
Increment Field SHIFT-key
The
INCR. FIELD
SHIFT-key is used to manually increase the field. This
is done at the defined ramp rate.
When the
INCR. FIELD
SHIFT-key is used, the current/field begins
ramping up. If the
INCR. FIELD
SHIFT-key is used again (while the
current/field is manually ramping up), the ramping will be paused.
Alternately, the
RAMP / PAUSE
key may be pressed to pause manual
ramping. Manual ramping will continue until paused or the Current
Limit
1
or Magnet Current Rating is achieved.
3.7.5
Field <> Current SHIFT-key
The
FIELD <> CURRENT
SHIFT-key is used to toggle between the use of
field units, either kG (kilogauss) or T (tesla), and the use of current units
(A)
2
. If the Model 430 Programmer is using field units (either kG or T) and
1. Refer to section 3.10.2.4 on page 54.
2. The value is always displayed in current (A) when an installed persistent switch is in
the
cooled state
since the value represents power supply current only, independent
of magnet current/field.
+40.92 A
Mode: Ramping
+2.50 Vs
V
PSwitch: ON
+0.00 A - Quench Detect @ +45.81 A
+0.00 Vs
PSwitch Heater: ON