Table 3-2, Ble 3-2 on, Operation – American Magnetics 05100PS-430-601 Integrated Power Supply System User Manual
Page 65
Rev. 5
47
Operation
Setup Menu : Supply
Note
The current must be less than 0.1% of I
max
in order to change the
Select Supply picklist value. If a change is attempted with current
above this value, the Model 430 Programmer will beep and ignore
Table 3-2.
Select Supply picklist values and associated parameters.
Power Supply
Min Output
Voltage (V)
Max Output
Voltage (V)
Min Output
Current (A)
Max Output
Current (A)
V-V Mode Input
Range (V)
AMI 08150PS
0
+8.000
+0.0000
+150.000
+0.000 to +10.000
AMI 12100PS
+0.000
+12.000
+0.000
+100.000
+0.000 to +10.000
AMI 12200PS
+200.000
AMI 4Q05100PS
−
5.000
+5.000
−
100.000
+100.000
−
10.000 to +10.000
AMI 4Q06125PS
-6.000
+6.000
-125.000
+125.000
AMI 4Q06250PS
-250.000
+250.000
AMI 4Q12125PS
-12.000
+12.000
-125.000
+125.000
AMI 10100PS
+0.000
+10.000
+0.000
+100.000
+0.000 to +5.000
AMI 10200PS
+200.000
AMI 05100PS
+10.000
a
+100.000
+0.000 to +10.000
AMI 05200PS
+200.000
AMI 05300PS
+300.000
AMI 05400PS
+400.000
AMI 05500PS
+500.000
HP 6260B
+10.000
+100.000
Kepco BOP 20-5M
b
−
10.000
−
5.000
+5.000
−
10.000 to +10.000
Kepco BOP 20-10M
−
10.000
+10.000
Xantrex XFR 7.5-140
+0.000
+7.500
+0.000
+140.000
+0.000 to +10.000
Custom...
c
−
20.000
+20.000
−
200.000
+200.000
−
10.000 to +10.000
a. The individual 05100PS power supply unit will source +10.000 Vdc at 100 A. How-
ever, the standard configuration of this series of power supplies includes the Model
601 Energy Absorber to provide bipolar operation. The 05x00-430-601 series of
power supply systems provides a maximum available voltage to the load of
±
5.000
Vdc at multiples of 100 A, up to 500 A depending on the system selected.
b. The Kepco BOP power supplies are limited to only one-half the output voltage
range since the supplies are designed to safely
dissipate
only one-half the rated
power output.
c. The values shown for the Custom... option are defaults. The user should enter the
appropriate values within the respective submenus. Custom values, once entered,
are saved in nonvolatile memory.