Ramp Configuration Commands and Queries
section 3.12 for additional information on determining ramp rates. Also
included are queries for collecting the magnet field, current, voltage, and
inductance.
•
CONFigure:VOLTage:LIMit <
voltage (V)
>
Sets the ramping Voltage Limit in volts. The ramping Voltage Limit may
not exceed the maximum output voltage of the power supply.
•
VOLTage:LIMit?
Returns the ramping Voltage Limit in volts.
•
CONFigure:CURRent:TARGet <
current (A)
>
Sets the target current in amperes.
•
CURRent:TARGet?
Returns the target current setting in amperes.
•
CONFigure:FIELD:TARGet <
field (kG, T)
>
Sets the target field in units of kilogauss or tesla, per the selected field
units. This command requires that a coil constant be defined, otherwise an
error is generated.
•
FIELD:TARGet?
Returns the target field setting in units of kilogauss or tesla, per the
selected field units. This query requires that a coil constant be defined,
otherwise an error is generated.
•
CONFigure:RAMP:RATE:SEGments <
value
>
Sets the number of ramp segments (see section 3.7.1 for details of the use
of ramp segments).
•
RAMP:RATE:SEGments?
Returns the number of ramp segments.
•
CONFigure:RAMP:RATE:CURRent <
segment>,
>
Sets the ramp rate for the specified segment (values of 1 through the
defined number of ramp segments are valid) in units of A/sec or A/min (per
the selected ramp rate units), and defines the current upper bound for that
segment in amperes (see section 3.7.1 for details of the use of ramp
segments).
•
RAMP:RATE:CURRent:<
segment>
?
Returns the ramp rate setting for the specified segment (values of 1
through the defined number of ramp segments are valid) in units of A/sec