5 model 601 energy absorber front panel layout, Figure 1-3. model 601 front panel layout, Figure 1-3 model 601 front panel layout – American Magnetics 05100PS-430-601 Integrated Power Supply System User Manual
Page 26: Introduction
8
Rev. 5
Introduction
Model 601 Front Panel
1.5 Model 601 Energy Absorber Front Panel Layout
The Fault LED is the only device on the Model 601 front panel. If the Fault
LED is not energized, the Model 601 is operating correctly. If the Fault
LED is energized, then one or more of the internal energy absorbing
elements has malfunctioned or power has been lost to the rear-panel
power connector. An audible alarm will also sound when the Fault LED is
energized.
Caution
If the system is in operation when an energy absorber fault occurs, a
safe magnet system state (typically zero current or a cooled
persistent switch in a connected magnet) should immediately be
attained. Do not continue to operate the unit, and refer to the
“Troubleshooting Hints” on page 137 for further direction.
Figure 1-3.
Model 601 Front Panel Layout