A.15.1 hardware and software requirements, A.15.2 preparation, In section a.15 on – American Magnetics 05100PS-430-601 Integrated Power Supply System User Manual
Page 206: Appendix
188
Rev. 5
Appendix
Model 430 Firmware Upgrade via Flash Card Reader
A.15 Upgrading the Model 430 Firmware via Flash Card Reader
Note
These instructions are intended primarily for a Model 430 being
upgraded from Version 1.59 or earlier. If the current version is v1.60
or later, upgrade should be performed via FTP according to
Note
These instructions apply specifically to the Windows XP
operating system. For other operating systems, please make
adjustments as appropriate.
This is a one-time only procedure - future upgrades will be handled via the
Ethernet communications connection on the rear panel of the instrument.
An operating system file on the Compact Flash (CF) card mounted inside
the Model 430 must be updated to enable the new web server
functionality. The file is not accessible via FTP which means the Model
430 cover must be removed, the CF card removed, and the files updated
using a CF card reader. Once this upgrade is completed, the CF card
should not have to be removed in order to perform future upgrades.
A.15.1
Hardware and Software Requirements
1.
The zip file, Model 430 flash card update.zip, is required for this
upgrade; the zipped files are to be extracted and copied to the CF
card.
2.
Host computer on a network to which the Model 430 can be
connected by standard Ethernet cable
or
Host computer that can be connected directly to the target Model
430 via an Ethernet “null-modem” or “crossover” cable.
3.
Standard Ethernet cable or Ethernet “null-modem” cable, as
appropriate.
A.15.2
Preparation
Complete the following in preparation for the upgrade.
1.
Power down and unplug the Model 430.
2.
Remove the cover from the Model 430 as follows:
a. Remove the screws securing the cover on the back edges of the
instrument (two near each side of the instrument).
b. Remove the screws securing the cover just behind the front panel (two
on each side of the instrument).
3.
Remove the CF card
1
from the Model 430 as follows: