beautypg.com

3 procedure for exiting persistent mode – American Magnetics 05100PS-430-601 Integrated Power Supply System User Manual

Page 101

background image

Rev. 5

83

Operation

Persistent Switch Control : Exiting Persistent Mode

Note

If desired, press the

ESC

key return the Model 430 Programmer to

the default display.

Note

Refer to section 3.6.2 on page 33 for the procedure to display the

magnet current that was established when the persistent switch was

cooled.

6. If desired, the power supply system can be de-energized. Turn the

power supply off first followed in a few seconds by the Model 430

Programmer.

Note

The Model 430 Programmer will store the state of the magnet in

memory and assist the user in exiting the persistent mode when the

Model 430 Programmer is next turned on (discussed in the section to

follow).

3.13.3 Procedure for Exiting Persistent Mode

To exit the persistent mode of magnet operation, the user should perform

the following steps:

1. If the Model 430 Programmer has not been powered off since the

magnet was placed in persistent mode, proceed to step 3.

2. If the Model 430 Programmer has been powered off since the

magnet was placed in persistent mode, complete the following

steps a. and b. before proceeding to step 3.

a. Energize the Model 430 and wait for the prompt on the Model

430 display and then energize the power supply.

After the power supply has been on for a few seconds, press

ENTER

to clear the Model 430 screen prompt.

+0.00 A -

Mode: Zero Current

+0.00 Vm

PSwitch Heater: OFF

+0.00 A

P

Turn on power supply

+0.00 Vs

Press ENTER to continue