AMI LIQUID HELIUM LEVEL SENSOR

INSTALLATION, OPERATION, AND

MAINTENANCE INSTRUCTIONS

I.

INTRODUCTION

The AMI liquid helium level sensor uses a small Niobium-Titanium (NbTi) wire as the
detector element. A heater creates and helps maintain a normal zone in that portion of
the wire above the liquid helium level while that portion of the wire below the liquid
helium level remains superconducting. The output voltage of the sensor varies linearly
with a change in liquid level.

The AMI liquid helium level sensor is designed to operate with an AMI liquid helium
level meter. Operation of the sensor with other level meters or operation of different
length sensor with a meter calibrated for a specific length may void the sensor
warranty.

II.

SPECIFICATIONS

Diameter:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1/4"

Active lengths:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 to 80 inches

Overall length: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . usually 1 inch longer than active

length (1/2 inch at top and bottom)

Sensor current: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75 milli-amperes (nominal)

Sensor voltage: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-60 V DC depending on sensor

length

Nominal sensor resistance:. . . . . . . . . . . . . . . . . . . . 4.5 ohms/cm (11.6 ohms/in.) @ 20K

5.4 ohms/cm (13.7 ohms/in.) @ 300K

Maximum magnetic field: . . . . . . . . . . . . . . . . . . . . . 10 Tesla

Caution symbol: necessary instructions in this document in order to protect
against damage to the product.