AMI LIQUID HELIUM LEVEL SENSOR
INSTALLATION, OPERATION, AND
MAINTENANCE INSTRUCTIONS
I.
INTRODUCTION
The AMI liquid helium level sensor uses a small Niobium-Titanium (NbTi) wire as the
detector element. A heater creates and helps maintain a normal zone in that portion of
the wire above the liquid helium level while that portion of the wire below the liquid
helium level remains superconducting. The output voltage of the sensor varies linearly
with a change in liquid level.
The AMI liquid helium level sensor is designed to operate with an AMI liquid helium
level meter. Operation of the sensor with other level meters or operation of different
length sensor with a meter calibrated for a specific length may void the sensor
warranty.
II.
SPECIFICATIONS
Diameter:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1/4"
Active lengths:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 to 80 inches
Overall length: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . usually 1 inch longer than active
length (1/2 inch at top and bottom)
Sensor current: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75 milli-amperes (nominal)
Sensor voltage: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-60 V DC depending on sensor
length
Nominal sensor resistance:. . . . . . . . . . . . . . . . . . . . 4.5 ohms/cm (11.6 ohms/in.) @ 20K
5.4 ohms/cm (13.7 ohms/in.) @ 300K
Maximum magnetic field: . . . . . . . . . . . . . . . . . . . . . 10 Tesla
Caution symbol: necessary instructions in this document in order to protect
against damage to the product.
Document Outline
- AMI LIQUID HELIUM LEVEL SENSOR INSTALLATION, OPERATION, AND MAINTENANCE INSTRUCTIONS
- I. INTRODUCTION
- II. SPECIFICATIONS
- III. INSTALLATION
- A. Carefully remove the sensor from the shipping tube and remove all packaging material.
- B. The sensor must be mounted with the electrical leads at the top.
- C. For minimum losses, mount the liquid helium sensor so that warm helium gas rising from the sen...
- D. The sensor may be mounted by taping or clipping it to an appropriate support structure. Do not...
- E. Avoid installing in a location where icing (frozen water or gas) may occur since ice formation...
- F. Ensure the level meter is de-energized (unplugged) and connect the sensor to the meter. The li...
- IV. OPERATION
- V. TROUBLESHOOTING
- VI. MAINTENANCE
- VII. WARRANTY
- VIII. RETURN AUTHORIZATION