2 target field setpoint key, 3 ramp / pause key, 4 ramp to zero key – American Magnetics 05100PS-430-601 Integrated Power Supply System User Manual

Page 51: Operation

background image

Rev. 5

33

Operation

Single-key Commands : Target Field Setpoint

3.6.2

Target Field Setpoint Key

Pressing the

TARGET FIELD SETPOINT

key provides a menu for setting

the target field/current. The target field/current is the field or current to

which the Model 430 Programmer ramps the superconducting magnet

when it is not paused. The target field/current may be set to the lesser of

(1) the Magnet Current Rating, and (2) the Current Limit

1

or equivalent

field (per defined coil constant). The target field/current does not require a

sign (attempting to enter a negative value will cause the Model 430

Programmer to produce one beep to indicate an error).

When on the default display, pressing the

SHIFT

+

TARGET FIELD

SETPOINT

keys will temporarily display the Magnet Current/Field for

three seconds before reverting to the default display. The value displayed

is as follows:

•When in driven mode, the present current/field will be displayed.
•When in persistent mode, the current/field will be displayed that

was flowing in the magnet at the time persistent switch was

cooled.

3.6.3

Ramp / Pause Key

Pressing the

RAMP / PAUSE

key toggles the Model 430 Programmer

between the ramping mode and the paused mode. If the

RAMP / PAUSE

key is pressed while the Model 430 is ramping, the ramping is paused. If

the

RAMP / PAUSE

key is pressed while the Model 430 is paused, the

Model 430 continues ramping.

3.6.4

Ramp To Zero Key

Pressing the

RAMP TO ZERO

key causes the Model 430 Programmer to

immediately begin ramping field/current up or down to zero field/current

at the defined ramp rate(s). Ramping to zero may be interrupted at any

time by pressing the

RAMP / PAUSE

key, which causes the Model 430

Programmer to enter the PAUSED mode and maintain the field/current

present at the point it was paused.

1. Refer to section 3.10.2.4 on page 54.

+50.00 A - Target Field (kG)

+0.50 Vs

+50.000

+0.25 A

P Magnet Current (A)

+0.00 Vs

+10.00 A