3 current leads energized indicator, 4 magnet quench indicator, 3 shift indicator – American Magnetics 05100PS-430-601 Integrated Power Supply System User Manual

Page 61: 9 setup menu, 1 entering / exiting setup menu, 2 menu navigation, Operation

background image

Rev. 5

43

Operation

LED Indicators : Current Leads Energized

The state of this LED is kept in nonvolatile memory when the

Model 430 is powered off, so that the LED state is retained even

during a power cycle of the Model 430. Thus, the

MAGNET IN

PERSISTENT MODE

LED is an indicator that the magnet is

persistent and has at least some persistent field.

3.8.2.3

Current Leads Energized Indicator

The blue

CURRENT LEADS ENERGIZED

LED indicates that at

least 100 mA of current is flowing in the Model 430 power supply

system output current leads.

3.8.2.4

Magnet Quench Indicator

The red

MAGNET QUENCH

LED indicates that a magnet quench

condition has been detected. See section 3.15 on page 88 for details.

3.8.3 SHIFT

Indicator

The green

SHIFT

LED indicates that the

SHIFT

key has been pressed, and

the next numeric keypad key pressed will actuate the shifted function

(shown in light blue) rather than the numeric keypad function. See

section 3.7 on page 34 for details of

SHIFT

key use.

3.9

Setup Menu

Setup of the Model 430 Programmer requires the user to navigate the

setup menu. Navigation of the setup menu is very intuitive — quite

similar, for example, to the use of a cell phone menu.

3.9.1

Entering / Exiting Setup Menu

To enter the setup menu, simply press the

MENU

key. When in any of the

setup menus, pressing the

MENU

key will exit the setup menu. The

MENU

key toggles the Model 430 Programmer in and out of setup mode.

Alternately, if the top level setup menu is being displayed, pressing the

ESC

key exits the setup menu.

3.9.2 Menu

Navigation

Pressing the MENU key enters the menu structure at the top level. The

display will look approximately as shown below:

+0.00 A - Setup Mode (Select one)

+0.50 Vs

Supply

Load

Misc