SECTION 8—SERVICE PROCEDURES

Part No 1114850

49

Recumbent Handcycle Series

Replacing the Crank Arms - XLT Gold, XLT PRO and
Top End Force

NOTE: For this procedure, refer to FIGURE 8.2 on page 50.

NOTE: Right and left determined from as if seated on the handcycle.

Replacing the Left Side Crank Arms

1. Remove the two socket screws from the existing crank arm clamp.

2. Remove socket screw and cap from the spindle.

3. Remove the existing crank arm from the spindle.

4. Position the new crank arm onto the spindle so that the new crank arm is evenly 

aligned with the right side crank arm.

5. Re install the socket screw and cap from the spindle. DO NOT over tighten.

6. Using the two socket screws, secure the new crank arm clamp to the spindle. Securely 

tighten.

Replacing the Right Side Crank Arms

1. Remove the two mounting bolts and locknuts securing the chain ring disk to the tabs 

of the existing crank arm.

2. Remove the two socket screws from the existing crank arm clamp.

3. Remove socket screw and cap from the spindle.

4. Remove the existing crank arm from the spindle.

5. Position the new crank arm onto the spindle.

6. Using the two mounting bolts and locknuts, secure the tabs of the new crank arm to 

the chain ring disk.

7. Ensure the new crank arm is evenly aligned with the left side crank arm.

8. Re install the socket screw and cap from the spindle. DO NOT over tighten.

9. Using the two socket screws, secure the new crank arm clamp to the spindle. Securely 

tighten.