Real time display mode – FEC DSP1500 (SAN3) User Manual
Page 72

PAGE 6-10
enFORCE
Operation Manual
6.3.4 Real Time Display Mode
The following data is displayed on the screen in the Real Time Display Mode.
The [D-NO] changes by pressing the [↑] and [↓] buttons.
0: Load
Value
Display
Displays present load applied to the load transducer in real time.
1: Load Voltage Value Display
Displays the present load signal voltage applied to the load transducer in real time.
2: Maximum Load Value Display
Displays the peak load measured from the last time the display was switched to this
mode. This is reset when leaving Real Time Display Mode.
This value can be held for 1 minute by pressing the [SET] button.
3: Distance Moved Display
Displays the absolute distance from the home position.
4: Electric Thermal Value
Displays the present electric thermal value.
5: Current
Value
Displays the present current value.
The maximum amplifier current value is 1023.
6: Resolver Position Data Display
Displays the position (in pulses) of the resolver at the present time.
7~: In-house Adjusting Display
Displays the in house adjustment value.
「
D-NO」 「DATA」
Load value Display
Load Voltage Value Display
Maximum Load Value Display
D-NO DATA D-NO
DATA
0
Load Value Display
4
Electric Thermal Value(Overheat when 100)
1
Load Voltage Value Display
5
Current Value( Max 1023 )
2
Maximum Load Value Display
(Maximum Value Hold)
6
Resolver Position Data Display(0~4093)
3
Distance Moved Display
7~ FEC
use
only
Chapter 6: System Operation
0
1
2
n
[↑],[↓]
Button
D-NO
: 1 figure